Liquid Chromatographic Determination of Dobutamine HCl in Pharmaceutical Formulations
K. Balamuralikrishna and B. Syamasundar*
Department of Chemistry, Acharya Nagarjuna University, Nagarjuna Nagar – 522 510 Andhra Pradesh, India
*Corresponding Author E-mail: profbsyamsundar@yahoo.co.in
ABSTRACT:
The objective of the study is to develop a reversed phase high performance liquid chromatographic method using ultra violet detection for the determination of Dobutamine HCl in bulk and pharmaceutical dosage forms. Phenomenex Luna C18 column (250 mm x 4.6 mm, 5 μ) was used as the stationary phase with a mixture of Acetonitrile:Methanol:Tetrahydrofuran (70:20:10 v/v/v) as the mobile phase. The response of the drug was linear in the concentration range of 10-30 µg/ml. Limit of detection and Limit of Quantification were found to be 0.34 and 1.1 µg/ml, respectively. The percentage recovery ranged between 99.54 and 100.06%. The factors affecting column separation of the analyte were studied. The results demonstrated that this method is reliable, reproducible and suitable for routine analysis.
KEYWORDS: Dobutamine HCl, High Performance Liquid Chromatography, Validation.
Dobutamine Hydrochloride (DOB)1 is chemically,(±)-4-(2-{[3-(p-Hydroxyphenyl)-1-methylpropyl] amino}ethyl) pyrocatechol hydrochloride, (Figure 1) and belongs to a series of pyrocatechol derivatives, which exhibit strong sympathomimetic with direct effects on beta1- adrenergic receptors, giving it a prominent inotropic action on the heart. It is used to the increase the contractility of the heart in acute heart failure, as occurs in cardiogenic shock and myocardial infarction.
Literature survey revealed that several analytical methods have been reported for the determination of Dobutamine hydrochloride in pure drug, pharmaceutical dosage forms and in biological samples using spectrophotometry2-15, Gas chromatography16 and liquid chromatography17-21 either in single or in combined forms. The retention time of the reported HPLC methods is more than five minutes and the flow rate of the mobile phase is 1.0 ml/min, which requires a greater mobile phase for the determination of Dobutamine HCl. Hence, in the present study it was intend to develop a rapid, economical, simple and precise RP-HPLC method for the estimation of Dobutamine HCl in bulk and formulations.
EXPERIMENTAL:
Chromatographic conditions:
The HPLC Agilent 1200 system was equipped with an auto injector (100 µl syringe ) and connected to a variable diode array wavelength detector and recorder. A Luna C18 reversed phased column (250 mm x 4.6 mm, 5 μ) bonded on to 5 µm silica gel, manufactured by Agilent, was used for analysis. A mixture of Acetonitrile: Methanol: Tetrahydrofuran (70:20:10 v/v/v, pH 6.5) was used as the mobile phase. The flow rate was 1.0 ml/min, wavelength was 292 nm, and the injection volume was 20 µl. The analysis was carried out at ambient temperature.
Reagents:
All the chemicals were of analytical grade and used without further purification. HPLC grade Acetonitrile (Chromasolv), Methanol (Qualigens), Tetrahydrofuran (Qualigens) and o-Phosphoric acid - ExcelaR grade (Qualigens) were used. The standard drug sample of the Dobutamine HCl was supplied by Sun pharmaceuticals Industries Limited, Mumbai.
Standard and sample preparation:
Standard stock solution at a concentration of 1 mg/1ml Dob was prepared by dissolving a 20 ml of methanol in 100 ml volumetric flask. This solution was suitably diluted with the mobile phase to get a working standard solution of 100 μg/mL of Dobutamine hydrochloride. Five injection vials of Dobutrex were taken. Their aluminium closures were removed. The powders in all the vials were pooled up and the average weight of powder in one vial was calculated. An amount of powder equivalent to 100 mg of Dobutamine hydrochloride was taken in a 100 mL volumetric flask and it was mixed with 20 mL of methanol. The contents of the flask were sonicated for about 20 min for complete solubility of the drug and the volume was made up to 100 mL with mobile phase to get a 1 mg/mL of the solution. Then the mixture was filtered through a 0.45 μ membrane filter. From the above solution a Ten mL aliquot was taken into a separate 100 mL volumetric flask and made up to the volume with mobile phase and mixed well to get a 20 μg/mL of the solution. The above solution (20 μL) was then injected eight times into the column.
Fig 1: Molecular structure of Dobutamine hydrochloride
Method validation:
The method validation parameters, such as, specificity, linearity, accuracy, precision, limit of detection, limit of quantification and robustness were verified as per the International conference on Harmonization (ICH) guidelines22.
Specificity:
The specificity of the method was confirmed by comparison of chromatograms obtained from the standard and the samples.
Linearity and range:
Linearity of the peak area response was determined by taking six replicate measurements at seven concentration points. Working dilutions of DOB in the range of 10-30 μg/mL were prepared by taking suitable dilutions of the standard solutions in different 10 mL volumetric flasks and diluted up to the mark with the mobile phase. Twenty microlitres of the dilution was injected six times into the column.
Accuracy (% Recovery):
The accuracy study was carried out by the standard addition method. Known amounts of standard solution of DOB (10,15 and 20 µg/ml) were added to a sample solution of DOB (20 µg/ml).Each solution was injected in triplicate and the percentage recovery was calculated.
Precision:
Standard solutions having concentrations of 15, 20 and 25µg/ml were prepared. For the intra-day studies three repeated measurements of these solutions were carried out within a day and %RSD was calculated. In the Inter-day variation studies, three reported measurements of the same solutions were carried out on three consecutive days and % RSD was calculated.
Limit of detection and Limit of quantification:
The limit of detection (LOD) and limit of quantification (LOQ) were calculated from the standard deviation of responses and slopes using signal-to-noise ratio as per ICH guidelines22.
Robustness:
Robustness of the method was studied by changing the extraction time of DOB, from dosage form by five minutes, composition of mobile phase by 2% of organic phase, flow rate by 0.1 ml/min, and wavelength by 1 nm.
RESULTS AND DISCUSSION:
Chromatographic conditions were developed empirically by selecting various solvents such as water, acetonitrile and methanol in different concentrations, and the retention time and resolution were assessed to get optimum conditions. Under the established chromatographic conditions, DOB was well-resolved. As the system suitability parameters such as tailing factor and theoretical plate numbers were within the acceptable limits, the resolution of peaks were considered optimum. Retention time for DOB for 5.008 minutes indicating the rapid analysis of Dobutamine [Figure 2]. The calibration curves were constructed by plotting the peak areas of DOB versus concentration of DOB, and the regression equation was calculated. Each response was the average of three determinations. The results [Table 1] obtained showed that the linearity range was 10-30µg/ml. The corresponding regression equation was found to be Y=9562.48 X +40.29 (Y is peak area of DOB and X is DOB concentration) with a correlation coefficient of 0.9999.
Table 1: Summary of regression analysis and validation parameters
|
Parameters |
Values |
|
Regression analysis Slope Intercept Correlation Validation parameters LOD (µg/ml) LOQ (µg/ml) Accuracy (%) ± % RSD Precision Intra-day (n=3) Inter-day (n=3) System suitability test parameters Retention time(min) ± % RSD Tailing factor ± % RSD Theoretical plates ± % RSD |
9562.48 40.29 0.9999
0.34 1.1 99.54 ± 1.2
0.38-0.83 0.38-0.72
5.008 ± 0.03 1.62 ± 0.1 3201± 1.45 |
The recovery study showed that the mean recovery was 99.54 % and the RSD for three samples was lower than 2.0%. The RSD values of intra-day and inter-day were found to be in the range of 0.38-0.83 and 0.38-0.72, respectively.
Fig 2: A typical Chromatogram showing the separation of Dobutamine hydrochloride Standard
Table 2: Recovery study data of DOB
|
Standard Dobutamine concentration (µg/ml) |
Sample Dobutamine concentration (µg/ml) |
Total amount of Dobutamine from standard graph (µg/ml) |
Recovery of standard drug (µg/ml) |
% Recovery of standard |
|
20 |
10 |
30.01 |
10.01 |
99.94 |
|
20 |
15 |
34.82 |
14.82 |
99.54 |
|
20 |
20 |
40.13 |
20.13 |
100.06 |
As these parameters were within the acceptable range, it was inferred that the method was precise. The accuracy of the developed method was assessed by the standard addition method and the RSD was found to be within the acceptable limit [Table 2]. The limit of quantification (LOQ) of the method, defined as the minimum concentration that could be measured, was found to be 1.1 µ/ml. The limit of detection (LOD) was 0.34 µg/ml [Table 1]. The method was found to be robust as the results were not significantly affected by the slight variation in the composition of mobile phase, flow rate and pH of the buffer. The chromatographic parameters such as selectivity and peak asymmetry are within the acceptable range. The number of theoretical plates and the tailing factor were observed to be satisfactory [Table 1].
Analysis of marketed formulation:
The proposed validated RP-HPLC method was successfully applied to determine DOB in two marketed formulations. The mean percentages of DOB were found to be 99.84 ± 0.92, 100.48 ± 1.01which were comparable to the corresponding labeled amounts [Table 3]. Furthermore, the matrix components, for example, the excipients, did not interfere with the analyte.
Table 3: Assay result for Dobutamine (mg of tablet) in the formulation product
|
Formulation |
Dobutamine in label claim (mg) |
Total amount of Dobutamine found (mg) |
% of Dobutamine found |
|
1 |
250 |
249.6 |
99.84 |
|
2 |
250 |
251.2 |
100.48 |
CONCLUSION
The validated RP-HPLC method employed here has proved to be rapid, simple, specific, accurate, precise, sensitive and robust. It can be successfully used for the routine analysis of DOB in bulk and in pharmaceutical dosage form.
REFERENCES
1. Martindale: The complete drug reference. 36th edition, Pharmaceutical press, Lambeth High Street, London. 1271-1272, 2009.
2. G. Ramana Rao, A. B. Avadhanulu and R. Giridhar. Spectrophotometric determination of dopamine hydrochloride, dobutamine hydrochloride in their dosage forms.Indian Drugs.25 (9):382-384(1988).
3. T. Guo-feng, H. Yu-ming, S. Wen- bing and L. Wei-bing. Determination of dobutamine hydrochloride by flow-injection Chemiluminescence method. Xinan shifan daxue xuebao, ziran kexueban. 29(4): 638-641(2004).
4. S. Fengmin, H.Shibin and H. Yuling. Flow-injection Chemiluminescence determination of dobutamine hydrochloride. Huaxue fenxijiliang. 12(4): 26-27(2003).
5. P. S. N. H. Ramachandra Rao, T. Siva Rao, U. Viplava Prasad and C. S. P. Sastry. Spectrophotometric methods for the determination of dobutamine and vancomycin in formulations. Indian pharmacist. 2(9): 59-61(2003).
6. EI-Kommos and E. Michael. Spectro photometric method for the determination of dobutamine hydrochloride. Analyst 108(1284):380-385(1983).
7. K. V. Kanna Rao, B. V. V. Ravi kumar, S. S. Rao and M. E. B Rao. Spectro photometric determination of salmeterol and dobutamine Gibb’s reagent. Acta Ciencia Indica 28(4):239-241(2002).
8. K. V. Kanna Rao and M. N. Reddy.Spectrophotometric determination of dobutamine hydrochloride in pharmaceutical formulations Asian Journal of Chemistry 13(4):1535-1538(2001).
9. S. P. Chilukuri Sastry, T. Siva Rao, S. N. H Pulipaka Ramachandra Rao and U.V. Prasad. Assay of vancomycin and dobutamine using sodium metaperiodate. Mikrochimica Acta 140(1-2):109-118(2002).
10. P. S. N. H. Ramachandra Rao, T. Siva Rao, U. Viplava Prasad and C. S. P. Sastry. Spectrophotometric determination of dobutamine and vancomycin using Fe (I I I). Journal of Institution of Chemists (India). 74(1): 25(2002).
11. T. Siva Rao, P. S. N. H. Ramachandra Rao, U. Viplava Prasad and C. S. P. Sastry. Spectrophotometric determination of vancomycin and dobutamine using ethyl acetoacetate.Journal of Institution of Chemists (India). 74(1):26-27(2002).
12. F. Shunli, W. Zhihao, Z. Lei, L. Chao and L. Jinming. Determination of dobutamine by flow injection Chemiluminescence method. Fenxi ceshi xuebao. 22(2):87-89(2003).
13. Z. Yan. Voltammetric behavior of dobutamine at poly-(Acridine orange) film midified electrode and its determination by adsorptive stripping voltammetry. Analytical letters, 37(10): 2031-2042(2004).
14. Q. Wang, J. Gendi, H. Xiaoya, L. Zongzhou and Y. Cheng. Determination of dobutamine with a platinum electrode in presence of large amount of ascarbic acid. Zhongguo yiyao gongye zazhi. 33(2): 87-90(2002).
15. G.J. Yang, J. J. Xu and H.Y. Chen. The study of redox mechanism of dobutamine at different pH media by electrochemical and in situ spectroelectrochemical methods. Electrochimica Acta. 49(19): 3121-3127(2004).
16. L. Jin, L.jianhua and C. Guiliang. Capillary gas chromatographic determination of organic residual in dobutamine hydrochloride using head space injection and direct injection of solution. Yaown fenxi zazhi. 26(1): 120-123(2006).
17. Y. Chen, C. Guiliang and Z. Hui. Determination of dobutamine hydrochloride and its related substances by RP- ion pair- HPLC. Zhongguo yiyao gongye zazhi. 32(7):307-309(2001).
18. D. Ross, H. Jane and C.w. William. Cardiotonic agents: a simple HPLC procedure for the quantitation of dobutamine and a new congener in plasma. Res. Commun. Chem. Pathol. Pharmacol. 48(2):313-316(1985).
19. E. H. Gregory and W. J. Lai. Determination of dobutamine in plasma by liquid chromatography with electrochemical detection. Anal. Lett. 16(B1): 69-75(1983).
20. G. Alberts, F. Boomsma, A. J. Man in t veld and M. A. D. H. Schalekamp. Simultaneous determination of catecholamines and dobutamine in human plasma and urine by High performance liquid chromatography with fluorimetric detection. J. Chromatogr. Biomed. Appl. 583(2): 236-40(1992).
21. Z. Yingxue, C. Funan and Z. Zhujun. Chemiluminescence determination of dobutamine hydrochloride in human serum by High performance liquid chromatographic separation connected with ultrafiltration technology. Fenxi Huaxue 32(6):769-771(2004).
22. ICH Harmonized Tripartite Guidelines (Q2R1). Validation of analytical. Validation of analytical procedures: Text and Methodology. International Conference on Harmonization, European commission, Japan and USA (2005).
Received on 08.11.2010 Modified on 20.11.2010
Accepted on 28.11.2010 © AJRC All right reserved
Asian J. Research Chem. 4(2): February 2011; Page 304-307